Dose Reproducibility in Axcelis GSD Implanters Using Stabil-Ion Gauge

Thursday, November 6, 2003, 11:00 am, Room 323

Session: Reproducibility, Precision, and Accuracy of Vacuum and Process Measurements
Presenter: R.C. Johnson, INNOViON Foundry Ion Implantation Engineering
Correspondent: Click to Email

Long-term dose reproducibility and tool to tool dose matching in the Axcelis GSD end-station is critically dependent on process chamber pressure measurement and Pressure Compensation factor selection. Pressure Compensation factor (PCOMP) determination is well-established. Pressure measurement in the GSD end-station depends on accurate, repeatable gauge capability: incorrect pressure measurements directly lead to dose errors. For example, the dose equation using PCOMP tells us that for a modest PCOMP value of 30%, a chamber pressure measurement error of 2E-5 torr can result in a dose error up to 6% at normal process pressures. The original HCIG used for pressure measurement was not capable of meeting the requirements for good dose control since gauge to gauge differences were not controlled and gauge accuracy was only on the order of 30%. Axcelis introduced the Granville-Phillips 360 Stabil-Ion gauge to improve dose reproducibility through much improved gauge to gauge matching (+/-6%) and more accurate gauge output. This paper discusses the details of the care and feeding of the Stabil-Ion gauge system and it’s impact on process dose and process trends.